Home / Recalls / Medical Devices
Bien-Air Surgery Sa recalls Bien-Air Surgery REF 1600407-001 SN **** FOOTCTRL OSSEODOC - Product Usage: intended to convert electrical energy to mec
Recalled October 28, 2020 · FDA recall Z-0176-2021 · Bien-Air Surgery Sa
Hazard
The magnet located inside the foot pedal may come off and may result in unattended start of unstoppable motor. · FDA Class II
Units
Total US = 1120 foot pedals
Sold at
US Nationwide distribution including in the states of CA, MI and MN.
Description
The magnet located inside the foot pedal may come off and may result in unattended start of unstoppable motor.
Product
Bien-Air Surgery REF 1600407-001 SN **** FOOTCTRL OSSEODOC - Product Usage: intended to convert electrical energy to mechanically drive handpieces and microsaws used to cut and shape bones.
Manufacturer(s)
Bien-Air Surgery Sa
Made in
Switzerland