Bio-Rad Laboratories, Inc. recalls EVOLIS Microplate System, Catalog # 89601. Part number 89788 for the EVOLIS Operator's Manual. in vitro diagnostic

Recalled February 24, 2016 · FDA recall Z-0807-2016 · Bio-Rad Laboratories

Hazard
The error "Washer Reagent Clean Fluid" level low alarmed, which prompted the customer to open the system drawer. When the system drawer was opened the customer noted the plate transport was still running and as a consequence the plate was pushed off the plate transporter. · FDA Class II
Units
278 units
Sold at
US Nationwide Distribution including Puerto Rico.

Description

The error "Washer Reagent Clean Fluid" level low alarmed, which prompted the customer to open the system drawer. When the system drawer was opened the customer noted the plate transport was still running and as a consequence the plate was pushed off the plate transporter.

Product
EVOLIS Microplate System, Catalog # 89601. Part number 89788 for the EVOLIS Operator's Manual. in vitro diagnostic Product Usage: EVOLIS Microplate System is a 4 plate fully integrated microplate processing system designed for use with multiple EIA assays.
Manufacturer(s)
Bio-Rad Laboratories, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov