Bio-Rad Labs recalls Autoimmune EIA Anti-Tissue Transglutaminase (tTG) IgA Test; Material Number 425-2260; Lot number 18200A - Product Usage:

Recalled May 6, 2020 · FDA recall Z-1800-2020 · Bio-Rad Labs

Hazard
An elevated lot-to-lot bias was observed for Autoimmune EIA Anti-Tissue Transglutaminase (tTG) IgA when compared to a previous lot of product. The observed elevated bias between these two lots is largely due to samples with higher levels of antibodies. With the cut-off for the assay being 15 U/mL, the clinical classifications for most samples would remain unchanged. · FDA Class II
Units
376 units
Sold at
US Nationwide distribution including the states of AZ, NY, LA, FL, CA, TX, IA, MS, UT, and IL.

Description

An elevated lot-to-lot bias was observed for Autoimmune EIA Anti-Tissue Transglutaminase (tTG) IgA when compared to a previous lot of product. The observed elevated bias between these two lots is largely due to samples with higher levels of antibodies. With the cut-off for the assay being 15 U/mL, the clinical classifications for most samples would remain unchanged.

Product
Autoimmune EIA Anti-Tissue Transglutaminase (tTG) IgA Test; Material Number 425-2260; Lot number 18200A - Product Usage: The assay is an aid in the diagnosis of celiac disease (gluten-sensitive enteropathy) and should be used in conjunction with other serological tests and clinical findings.
Manufacturer(s)
Bio-Rad Labs

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov