Home / Recalls / Medical Devices
Biocare Medical, LLC recalls 4plus Streptavidin HRP Label, Conjugated Streptavidin Horseradish Peroxidase, Detection Component, Catalog Number: HP60
Recalled June 17, 2020 · FDA recall Z-2291-2020 · Biocare Medical
Hazard
Firm has identified product has no staining. If used, may result in invalid test results when used with controls in the clinical setting. Positive control would observe no staining. The clinical impact may result in the consumption of patient tissue sample and a delay in diagnosis. · FDA Class II
Units
8
Sold at
US: TX, MA, TN, MO, IN
Description
Firm has identified product has no staining. If used, may result in invalid test results when used with controls in the clinical setting. Positive control would observe no staining. The clinical impact may result in the consumption of patient tissue sample and a delay in diagnosis.
Product
4plus Streptavidin HRP Label, Conjugated Streptavidin Horseradish Peroxidase, Detection Component, Catalog Number: HP604H
Manufacturer(s)
Biocare Medical, LLC
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →