RecallWatchUSA

Home / Recalls / Medical Devices

Biocare Medical, LLC recalls BIOCARE Medical , IntelliPath FLX, Automated Staining Instrument, Ref: " Model Number:IPS0001US and IPS0001INTL contain

Recalled April 7, 2021 · FDA recall Z-1333-2021 · Biocare Medical

Hazard
Their is a potential that the automated staining instrument with software version 3.5.3.1 may move with random speeds in random direction across its range of motion. This could result in the instrument hitting end stops or running into other components on the working deck. This could also result in wash buffer being dispensed on random locations. · FDA Class II
Units
72 units
Sold at
US: AL, AZ, CA, CT, FL, IL, MA, MD, MN, NC, OH, OR, PA, WA, WY OUS: Germany, Netherlands

Description

Their is a potential that the automated staining instrument with software version 3.5.3.1 may move with random speeds in random direction across its range of motion. This could result in the instrument hitting end stops or running into other components on the working deck. This could also result in wash buffer being dispensed on random locations.

Product
BIOCARE Medical , IntelliPath FLX, Automated Staining Instrument, Ref: " Model Number:IPS0001US and IPS0001INTL containing software version 3.5.3.1
Manufacturer(s)
Biocare Medical, LLC

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov