Home / Recalls / Medical Devices
Biomet, Inc. recalls Central Screw 6.5x25mm - Product Usage: used as a sterile screw that is intended for implantable shoulder prosthesis ass
Recalled February 24, 2021 · FDA recall Z-1131-2021 · Biomet
Hazard
Zimmer Biomet is conducting a lot specific medical device recall for the Comprehensive Reverse Shoulder Central Screw, 25 MM Length. It has been determined through a complaint investigation that a 40 MM length central screw was packaged and labeled as a 25 MM length central screw. · FDA Class II
Units
49
Sold at
US Nationwide distribution including in the states of MS, MN, KY, FL, MI, OH, AZ, IL, WI, CA, UT, TN.
Description
Zimmer Biomet is conducting a lot specific medical device recall for the Comprehensive Reverse Shoulder Central Screw, 25 MM Length. It has been determined through a complaint investigation that a 40 MM length central screw was packaged and labeled as a 25 MM length central screw.
Product
Central Screw 6.5x25mm - Product Usage: used as a sterile screw that is intended for implantable shoulder prosthesis assembly and/or fixation.
Manufacturer(s)
Biomet, Inc.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →