Home / Recalls / Medical Devices
Biomet, Inc. recalls HipLOC CHS Plate/Lag Screw Introducer
Recalled March 2, 2016 · FDA recall Z-0885-2016 · Biomet
Hazard
A complaint was received where the Hiploc Plate/Lag Screw Introducer was undersized and would not accept the introducor coupling screw. It appears now that not all parts manufactured to Rev A drawing were reworked. · FDA Class II
Units
23 units
Sold at
US Distribution to the states of : MO, UT, AZ, IL, PA, KY, MI, NC, NJ, MN and GA., and Internationally to The Netherlands.
Description
A complaint was received where the Hiploc Plate/Lag Screw Introducer was undersized and would not accept the introducor coupling screw. It appears now that not all parts manufactured to Rev A drawing were reworked.
Product
HipLOC CHS Plate/Lag Screw Introducer
Manufacturer(s)
Biomet, Inc.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →