RecallWatchUSA

Home / Recalls / Medical Devices

Biomet, Inc. recalls OSS Reamer Sleeve REF132-472661 KNEE REAMER SLEEVE 8.0 THRU 12.5 STAINLESS STEEL The OSS Knee Reamer Sleeves are prim

Recalled December 12, 2012 · FDA recall Z-0471-2013 · Biomet

Hazard
Biomet has initiated this action following an investigation which identified that the thin shaft reamer may get stuck and not slide into the sleeve. · FDA Class II
Units
46
Sold at
Worldwide distribution: USA (nationwide) and countries of: Japan and Netherlands.

Description

Biomet has initiated this action following an investigation which identified that the thin shaft reamer may get stuck and not slide into the sleeve.

Product
OSS Reamer Sleeve REF132-472661 KNEE REAMER SLEEVE 8.0 THRU 12.5 STAINLESS STEEL The OSS Knee Reamer Sleeves are primarily used to guide the bullet tip of cylindrical reamers during the tibial preparation in OSS procedures. They are reusable.
Manufacturer(s)
Biomet, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Biomet, Inc. recalls OSS Reamer Sleeve REF132-472661 KNEE REAMER SLEEVE 8.0 THRU 12.5 STAINLESS STEEL The OSS Knee Reamer Sleeves are prim | RecallWatchUSA