RecallWatchUSA

Home / Recalls / Medical Devices

Biomet, Inc. recalls PHOENIX, Interlocking screw for IM nail, Item numbers 14-405040, 14-405042, 14-405044, 14-405070 - Product Usage: Fractu

Recalled December 23, 2020 · FDA recall Z-0625-2021 · Biomet

Hazard
Specific lots of multiple sterile medical devices may have a breach in the sterile seal. A breach in the packaging seal could lead to loss of sterility of the device. · FDA Class II
Units
7428
Sold at
Worldwide distribution. US nationwide, ARGENTINA, AUSTRALIA, Brazil, CANADA, CHILE, CHINA, COSTA RICA, EMEA, HONG KONG, INDIA, JAPAN, KOREA, MALAYSIA, NEW ZEALAND, SINGAPORE, and TAIWAN

Description

Specific lots of multiple sterile medical devices may have a breach in the sterile seal. A breach in the packaging seal could lead to loss of sterility of the device.

Product
PHOENIX, Interlocking screw for IM nail, Item numbers 14-405040, 14-405042, 14-405044, 14-405070 - Product Usage: Fracture fixation of small bones, small bone fragements and long bones.
Manufacturer(s)
Biomet, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov