Biomet, Inc. recalls REF S313152 Straight Acetabular Inserter Handle Ringloc non-sterile Orthopedic manual surgical instrument
Recalled March 20, 2013 · FDA recall Z-0925-2013 · Biomet
Hazard
The Straight Ringloc Button Latch Inserter Handle may have been manufactured at the higher end of the tolerance causing interference with the Ringloc cup. The inserter handle may lock onto the cup during impaction. If alternative inserters are not available this could cause a surgical delay greater than 30 minutes. · FDA Class II
Units
58
Sold at
Worldwide Distribution - No USA distribution -Internationally to Canada, Japan, and Chile
Description
The Straight Ringloc Button Latch Inserter Handle may have been manufactured at the higher end of the tolerance causing interference with the Ringloc cup. The inserter handle may lock onto the cup during impaction. If alternative inserters are not available this could cause a surgical delay greater than 30 minutes.
Product
REF S313152 Straight Acetabular Inserter Handle Ringloc non-sterile Orthopedic manual surgical instrument
Manufacturer(s)
Biomet, Inc.