Biomet, Inc. recalls Superior Approach Resection Guide Assembly. Indicated for primary, fracture, or revision total shoulder replacement f
Recalled June 26, 2013 · FDA recall Z-1548-2013 · Biomet
Hazard
Biomet has initiated this action following an investigation which identified that the slot on the resection guide is offset in the wrong direction. When the slot on the resection guide is offset in the wrong direction, the Steinman pin that is intended to hold the guide into place may not come into contact with the bone. · FDA Class II
Units
17
Sold at
Worldwide Distribution-USA (nationwide) including the states of UT, TX, AL, KY, and WI, and the countries of Australia, The Netherlands, Korea, and Canada.
Description
Biomet has initiated this action following an investigation which identified that the slot on the resection guide is offset in the wrong direction. When the slot on the resection guide is offset in the wrong direction, the Steinman pin that is intended to hold the guide into place may not come into contact with the bone.
Product
Superior Approach Resection Guide Assembly. Indicated for primary, fracture, or revision total shoulder replacement for the relief of pain.
Manufacturer(s)
Biomet, Inc.