Home / Recalls / Medical Devices
Boston Scientific Corporation recalls ACCOLADE DR SL MRI (Model L311)
Recalled October 22, 2025 · FDA recall Z-0082-2026 · Boston Scientific
Hazard
Software to enhance Safety Architecture is available and designed to prevent initiation of Safety Mode in an ambulatory setting due to a high battery impedance state for the ACCOLADE family of devices which includes ACCOLADE, PROPONENT, ESSENTIO, and ALTRUA 2 dual chamber (DR) standard life (SL) and DR extended life (EL) pacemakers; and VISIONIST and VALITUDE cardiac resynchronization therapy pacemakers (CRT-Ps). · FDA Class I
Units
90,931 units
Sold at
Worldwide
Description
Software to enhance Safety Architecture is available and designed to prevent initiation of Safety Mode in an ambulatory setting due to a high battery impedance state for the ACCOLADE family of devices which includes ACCOLADE, PROPONENT, ESSENTIO, and ALTRUA 2 dual chamber (DR) standard life (SL) and DR extended life (EL) pacemakers; and VISIONIST and VALITUDE cardiac resynchronization therapy pacemakers (CRT-Ps).
Product
ACCOLADE DR SL MRI (Model L311)
Manufacturer(s)
Boston Scientific Corporation
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →