Home / Recalls / Medical Devices
Boston Scientific Corporation recalls Altrua 2 pacemaker, Models: a) S701 b) S702 c) S722
Recalled July 14, 2021 · FDA recall Z-2001-2021 · Boston Scientific
Hazard
Latent release of small amounts of hydrogen within the pacemaker may cause a low voltage capacitor to become electrically compromised over time resulting in accelerated battery depletion of the battery and associated progression of displayed battery depletion indicators. · FDA Class II
Units
2952 units
Sold at
Worldwide distribution.
Description
Latent release of small amounts of hydrogen within the pacemaker may cause a low voltage capacitor to become electrically compromised over time resulting in accelerated battery depletion of the battery and associated progression of displayed battery depletion indicators.
Product
Altrua 2 pacemaker, Models: a) S701 b) S702 c) S722
Manufacturer(s)
Boston Scientific Corporation
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →