RecallWatchUSA

Home / Recalls / Medical Devices

Boston Scientific Corporation recalls Boston Scientific Imager II Angiographic Catheter, M001316581 (5-pack outer package UPN), M001316580, single unit, inne

Recalled April 19, 2017 · FDA recall Z-1781-2017 · Boston Scientific

Hazard
One lot of Imager II Angiographic Catheters is being recalled due to packaging mislabeling. The product may be labeled as Imager II Contra 2 curve catheters, when in fact the package contains Contra curve catheters. · FDA Class II
Units
26 boxes of 5 (130 singles)
Sold at
AZ, FL, MO, NJ, NY, OH, PA, and TX.

Description

One lot of Imager II Angiographic Catheters is being recalled due to packaging mislabeling. The product may be labeled as Imager II Contra 2 curve catheters, when in fact the package contains Contra curve catheters.

Product
Boston Scientific Imager II Angiographic Catheter, M001316581 (5-pack outer package UPN), M001316580, single unit, inner package UPN)
Manufacturer(s)
Boston Scientific Corporation