Home / Recalls / Medical Devices
Boston Scientific Corporation recalls Flexiva Pulse 242 TracTip Single-Use Laser Fibers - intended to be used as a device that transmits Ho:YAG laser energy f
Recalled August 2, 2023 · FDA recall Z-2229-2023 · Boston Scientific
Hazard
Manufactured with the incorrect component may result in lower power output and degradation or dimming of the aiming beam, leading to reduced efficiency of fiber performance. Continued use may cause overheating of the fiber connector, which could result in a burn, if touched · FDA Class II
Units
339 units
Sold at
Worldwide distribution - US Nationwide and the country of Canada.
Description
Manufactured with the incorrect component may result in lower power output and degradation or dimming of the aiming beam, leading to reduced efficiency of fiber performance. Continued use may cause overheating of the fiber connector, which could result in a burn, if touched
Product
Flexiva Pulse 242 TracTip Single-Use Laser Fibers - intended to be used as a device that transmits Ho:YAG laser energy from cleared laser consoles to urological anatomy Single-pack UPN : M006L8405960
Manufacturer(s)
Boston Scientific Corporation
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →