RecallWatchUSA

Home / Recalls / Medical Devices

Boston Scientific Corporation recalls PROPONENT DR Pacemaker, SL (Model Number L201), (VDD) SL (Model Number L209), SL MRI (Model Number L211), EL (Model Numb

Recalled January 22, 2025 · FDA recall Z-0823-2025 · Boston Scientific

Hazard
A subset of devices from the ACCOLADE Family of Pacemakers and CRT-Ps, built before September 2018, have an increased potential to initiate Safety Mode during telemetry or other normal, higher-power operations due to latent high battery impedance. · FDA Class I
Units
156,329 units (59,937 US, 96,792 OUS) in total
Sold at
US (Nationwide) including Puerto Rico and US Virgin Islands and OUS (foreign) to countries of: Guam and Northern Mariana Islands.

Description

A subset of devices from the ACCOLADE Family of Pacemakers and CRT-Ps, built before September 2018, have an increased potential to initiate Safety Mode during telemetry or other normal, higher-power operations due to latent high battery impedance.

Product
PROPONENT DR Pacemaker, SL (Model Number L201), (VDD) SL (Model Number L209), SL MRI (Model Number L211), EL (Model Number L221), and EL MRI (Model Number L231).
Manufacturer(s)
Boston Scientific Corporation

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov