RecallWatchUSA

Home / Recalls / Medical Devices

Boston Scientific Corporation recalls VALITUDE CRT-P pacemaker, Models: a) U125 b) U128

Recalled July 14, 2021 · FDA recall Z-2002-2021 · Boston Scientific

Hazard
Latent release of small amounts of hydrogen within the pacemaker may cause a low voltage capacitor to become electrically compromised over time resulting in accelerated battery depletion of the battery and associated progression of displayed battery depletion indicators. · FDA Class II
Units
6377 units
Sold at
Worldwide distribution.

Description

Latent release of small amounts of hydrogen within the pacemaker may cause a low voltage capacitor to become electrically compromised over time resulting in accelerated battery depletion of the battery and associated progression of displayed battery depletion indicators.

Product
VALITUDE CRT-P pacemaker, Models: a) U125 b) U128
Manufacturer(s)
Boston Scientific Corporation

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Boston Scientific Corporation recalls VALITUDE CRT-P pacemaker, Models: a) U125 b) U128 | RecallWatchUSA