RecallWatchUSA

Home / Recalls / Medical Devices

Boston Scientific Neuromodulation Corporation recalls Vercise Deep Brain Stimulation Systems Surgical Implant Manual, Document Number: 92328632-09

Recalled September 24, 2025 · FDA recall Z-2558-2025 · Boston Scientific Neuromodulation

Hazard
Wire break(s) have occurred in rechargeable deep brain stimulation implantable pulse generators, implanted in the submuscular pectoral location. Wire breaks prevent successful stimulation, thus requiring device removal/replacement and users may experience high monopolar impedances, undesired sensation, loss of therapy, return of symptoms and/or Bluetooth connectivity challenges. · FDA Class II
Units
25,260
Sold at
Worldwide - US Nationwide distribution including in the states of MN, FL, CO, MD, NY, NE, PA, WI, AZ, TN, MO, TX, MI, MT, MA, WV, AL, VA, NC, CA, OR, WA, LA, OH, NJ, NH, SD, GA, CT, NV, UT, IN, ID, IL, ME, SC, DC, OK, KY, AK, HI, RI, AR, KS, NM and the countries of Argentina, Brazil, Chile, Ecuador, Colombia , Mexico, Canada, Japan, Algeria, Austria, Azerbaijan, Belgium, Bulgaria, Croatia, Denmark, Egypt, Finland, France, Georgia, Germany, Greece, Hungary, Iceland, Iran, Ireland, Israel, Italy, Jordan, Kazakhstan, Kuwait, Lebanon, Libya, Luxembourg, Malta, Morocco, Netherlands, Norway, Pakistan, Poland, Portugal, Romania, Russian Fed., Saudi Arabia, Serbia, Slovakia, South Africa, Spain, Sweden, Switzerland, Turkiye, UAE, Ukraine, United Kingdom, Hong Kong, China, Singapore, Korea, Taiwan , India, Thailand.

Description

Wire break(s) have occurred in rechargeable deep brain stimulation implantable pulse generators, implanted in the submuscular pectoral location. Wire breaks prevent successful stimulation, thus requiring device removal/replacement and users may experience high monopolar impedances, undesired sensation, loss of therapy, return of symptoms and/or Bluetooth connectivity challenges.

Product
Vercise Deep Brain Stimulation Systems Surgical Implant Manual, Document Number: 92328632-09
Manufacturer(s)
Boston Scientific Neuromodulation Corporation

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Boston Scientific Neuromodulation Corporation recalls Vercise Deep Brain Stimulation Systems Surgical Implant Manual, Document Number: 92328632-09 | RecallWatchUSA