RecallWatchUSA

Home / Recalls / Drugs

Bound Tree Medical, LLC recalls Curaplex Epi Safe Kit, 8600-01102 TRAINING ONLY, Rx Only, Distributed by Sarnova HC, LLC s family of companies: Bound Tr

Recalled January 30, 2019 · FDA recall D-0385-2019 · Bound Tree Medical

Hazard
Labeling Incorrect Instructions: This recall has been initiated because the Instructions for Use included in the Epi-Safe Kits for the Epi-Safe Syringe (and/or the Safety Lok Syringe in Model 8600-01120) recommend a midpoint dosage of epinephrine for children between 0 lbs. and 66 lbs. of 0.15mL that has not been approved by FDA for children. Reliance on the current IFU and/or use of the Epi-Safe Syringe as instructed may result in the administration of an inappropriate dosage of epinephrine to young children experiencing anaphylaxis, especially those under 33 lbs. Additionally the Epi Safe syringe does not have the correct markings to allow a health care provider to draw up the correct dosage for children under 33 kg. · FDA Class II
Units
2,654 kits (6 syringes per kit / 1 IFU per kit)
Sold at
Nationwide in the USA

Description

Labeling Incorrect Instructions: This recall has been initiated because the Instructions for Use included in the Epi-Safe Kits for the Epi-Safe Syringe (and/or the Safety Lok Syringe in Model 8600-01120) recommend a midpoint dosage of epinephrine for children between 0 lbs. and 66 lbs. of 0.15mL that has not been approved by FDA for children. Reliance on the current IFU and/or use of the Epi-Safe Syringe as instructed may result in the administration of an inappropriate dosage of epinephrine to young children experiencing anaphylaxis, especially those under 33 lbs. Additionally the Epi Safe syringe does not have the correct markings to allow a health care provider to draw up the correct dosage for children under 33 kg.

Product
Curaplex Epi Safe Kit, 8600-01102 TRAINING ONLY, Rx Only, Distributed by Sarnova HC, LLC s family of companies: Bound Tree Medical, LLC, Cardio Partners, Inc., Emergency Medical Products, Inc. & Tri-anim Health Services, Inc. 5000 Tuttle Crossing Blvd. Dublin, OH 43016
Manufacturer(s)
Bound Tree Medical, LLC

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Bound Tree Medical, LLC recalls Curaplex Epi Safe Kit, 8600-01102 TRAINING ONLY, Rx Only, Distributed by Sarnova HC, LLC s family of companies: Bound Tr | RecallWatchUSA