Bovie Medical Corporation recalls BVX-330NR Bovie Precise 360 Handpiece 33cm, needle Product is used for the delivery of helium gas plasma for cuttin
Recalled August 31, 2016 · FDA recall Z-2647-2016 · Bovie Medical
Hazard
The adhesive may be inadequately applied or the assembly may be manipulated after application of the adhesive resulting in an incomplete bond to the angled tip. · FDA Class II
Units
95
Sold at
Worldwide Distribution - US Nationwide in the states of: MD, GA, MO and FL and the country of Finland.
Description
The adhesive may be inadequately applied or the assembly may be manipulated after application of the adhesive resulting in an incomplete bond to the angled tip.
Product
BVX-330NR Bovie Precise 360 Handpiece 33cm, needle Product is used for the delivery of helium gas plasma for cutting, coagulation, and ablation of soft tissue during open and laparoscopic surgical procedures.