RecallWatchUSA

Home / Recalls / Medical Devices

Bunnell, Inc. recalls Patient Circuit used with the Life Pulse High Frequency Ventilator. The Patient Circuit is a required disposable compon

Recalled December 26, 2012 · FDA recall Z-0514-2013 · Bunnell

Hazard
Customer complaints received indicate the heater wire insulation can melt, causing a short which creates sparking and smoke, in the circuit close to the humidifier cartridge. Bunnell is recalling all lots of circuits distributed between March 12, 2012 and November 30, 2012. · FDA Class I
Units
5743 patient circuits distributed in commerce
Sold at
Worldwide Distribution-USA (nationwide) and the countries of Canada, Australia, Malaysia, and Uruguay (animal study only).

Description

Customer complaints received indicate the heater wire insulation can melt, causing a short which creates sparking and smoke, in the circuit close to the humidifier cartridge. Bunnell is recalling all lots of circuits distributed between March 12, 2012 and November 30, 2012.

Product
Patient Circuit used with the Life Pulse High Frequency Ventilator. The Patient Circuit is a required disposable component of the Life Pulse High Frequency Ventilator. Model #(s): Individual Patient Circuit - catalog # 902 Patient Circuit Kit - catalog # 937: contains 2 circuits and 2 x 2.5 & 2 x 3.5 LifePort Adapters. Labeling states manufactured by: Bunnell Incorporated 436 Lawndale Drive, Salt Lake City, Utah 84115. Used for ventilating critically ill infants with pulmonary interstitial emphysema and infants with respiratory distress syndrome complicated by pulmonary air leaks, who are, in the opinion of their physicians, failing on conventional ventilation. The Patient Circuit is to provide a conduit for the humidification, warming, and temperature monitoring of the pressurized gas. The Patient Circuit is indicated for seven day single use.
Manufacturer(s)
Bunnell, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Bunnell, Inc. recalls Patient Circuit used with the Life Pulse High Frequency Ventilator. The Patient Circuit is a required disposable compon | RecallWatchUSA