Home / Recalls / Medical Devices
Canon Medical System, USA, INC. recalls Canon Aplio ultrasound system, Model: Aplio i900 TUS-AI900, Aplio i800 TUS-AI800, Aplio i700 TUS-AI700, Aplio a550 CUS-A
Recalled November 16, 2022 · FDA recall Z-0198-2023 · Canon Medical System, USA
Hazard
After completion, ultrasound system's Stress Echo may not terminate if user does not manually terminate the protocol. During next patient exam, an abnormal confirmation message will be displayed "Same view was saved in this phase. Do you use this image with StressEcho Review?" If "no", acquired images not saved, and if "yes" images assigned to the previous patient; could cause treatment delay. · FDA Class II
Units
65
Sold at
US: MT, MO, NC, SC, MN, WA, KY, TN, AR, IL, IA, MI, WV, KS, CA, HA, CO, LA, OK, ME, KY, NY, PA, GA, FL
Description
After completion, ultrasound system's Stress Echo may not terminate if user does not manually terminate the protocol. During next patient exam, an abnormal confirmation message will be displayed "Same view was saved in this phase. Do you use this image with StressEcho Review?" If "no", acquired images not saved, and if "yes" images assigned to the previous patient; could cause treatment delay.
Product
Canon Aplio ultrasound system, Model: Aplio i900 TUS-AI900, Aplio i800 TUS-AI800, Aplio i700 TUS-AI700, Aplio a550 CUS-AA550, Aplio a450 CUS-AA450
Manufacturer(s)
Canon Medical System, USA, INC.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →