Canon Medical System, USA, INC. recalls Canon Aquilion Precision, Model TSX-304A

Recalled April 22, 2020 · FDA recall Z-1725-2020 · Canon Medical System, USA

Hazard
Scanning may be interrupted due to an error during execution of the eXam Plan and the system is unable to be shut down normally, resulting in the need to forcibly turn the system power off and reboot. Loss of the acquired raw data would occur. · FDA Class II
Units
1 device
Sold at
Distribution was made to AK, AR, CA, FL, IA, ID, IL, KS, MA, MD, MI, MN, MO, MT, NC, ND, NH, NY, OH, OK, OR, PA, TN, TX, VT, WA, WI, and WV. There was government and military distribution but no foreign distribution.

Description

Scanning may be interrupted due to an error during execution of the eXam Plan and the system is unable to be shut down normally, resulting in the need to forcibly turn the system power off and reboot. Loss of the acquired raw data would occur.

Product
Canon Aquilion Precision, Model TSX-304A
Manufacturer(s)
Canon Medical System, USA, INC.