Canon Medical System, USA, INC. recalls Flat Panel Detector, Model Number: TFP-800A, used with the Alphenix Systems, Item Numbers: INFX-8000V, INFX-8000C, INFX-

Recalled January 19, 2022 · FDA recall Z-0492-2022 · Canon Medical System, USA

Hazard
Abnormal images such as horizontal stripes may be displayed and recovery may not not be possible by restarting the system. The worst case scenario is an operator who is working without a workable image may remove a catheter from the patient's blood vessel and damage the blood vessel. · FDA Class II
Units
8
Sold at
US Nationwide distribution in the states of DE, FL, LA, MT, NJ, OH, and WV.

Description

Abnormal images such as horizontal stripes may be displayed and recovery may not not be possible by restarting the system. The worst case scenario is an operator who is working without a workable image may remove a catheter from the patient's blood vessel and damage the blood vessel.

Product
Flat Panel Detector, Model Number: TFP-800A, used with the Alphenix Systems, Item Numbers: INFX-8000V, INFX-8000C, INFX-8000F, and INFX-8000H.
Manufacturer(s)
Canon Medical System, USA, INC.