Canon Medical System, USA, INC. recalls Infinix DP- i(Infinix-8000F) CAT-870B Catheterization Table (Titling) - Product Usage: The catheterization table is used

Recalled November 20, 2019 · FDA recall Z-0360-2020 · Canon Medical System, USA

Hazard
Table tilts without operator involvement as a result of liquid spilling into tilt console triggering a short circuit. · FDA Class II
Units
9 systems
Sold at
Worldwide distribution - US Nationwide distribution in the states of AL, CT, FL, IL, OR, PA, and TX; and country of Netherlands.

Description

Table tilts without operator involvement as a result of liquid spilling into tilt console triggering a short circuit.

Product
Infinix DP- i(Infinix-8000F) CAT-870B Catheterization Table (Titling) - Product Usage: The catheterization table is used with Infinix-8000F System (Interventional Angiography System), to position patient during diagnostic and interventional x-ray angiography.
Manufacturer(s)
Canon Medical System, USA, INC.