RecallWatchUSA

Home / Recalls / Medical Devices

Canon Medical System, USA, INC. recalls VANTAGE TITAN 3T Model MRT-3010/MEXL-3010

Recalled January 21, 2026 · FDA recall Z-1018-2026 · Canon Medical System, USA

Hazard
There is a potential for formation of ice occurring in the venting system of the superconducting magnet of the MRI systems. if a quench occurs in this condition, helium gas may not be able to escape via the designed vent paths, and the pressure in the helium vessel may rise. This pressure build-up could eventually rupture the helium vessel, causing helium gas to be released into the MR scanning room. · FDA Class II
Units
22 systems
Sold at
U.S. Nationwide distribution in the states of CA, CO, FL, IA, IL, IN, KS, MD, MO, MT, NV, NY, OH, OR, PA, PR, TX, VA, and WI.

Description

There is a potential for formation of ice occurring in the venting system of the superconducting magnet of the MRI systems. if a quench occurs in this condition, helium gas may not be able to escape via the designed vent paths, and the pressure in the helium vessel may rise. This pressure build-up could eventually rupture the helium vessel, causing helium gas to be released into the MR scanning room.

Product
VANTAGE TITAN 3T Model MRT-3010/MEXL-3010
Manufacturer(s)
Canon Medical System, USA, INC.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov