Capintec Inc recalls Captus 4000e Thyroid Uptake System, Model Numbers 5430-30151 and 5430-30152 - Product Usage: is intended to be used by t

Recalled April 1, 2020 · FDA recall Z-1564-2020 · Capintec

Hazard
It is possible that the collimators have screw hole locations that were insufficiently tapped. There is a potential for the collimator assembly to fall and come into contact with a patient or operator. · FDA Class II
Units
48 (481 additional units added 3/26/20)
Sold at
Worldwide distribution. US Nationwide, Australia, Austria, Bangladesh, Brazil, Canada, Chile, Colombia, Costa Rica, Ecuador, France, Hong Kong, India, Israel, Italy, Korea, Malaysia, Pakistan, Philippines, Poland, Saudi Arabia, Spain, Sweden, Taiwan, Thailand, Turkey, UAE, UK, and Vietnam. (list updated 3/26/2020)

Description

It is possible that the collimators have screw hole locations that were insufficiently tapped. There is a potential for the collimator assembly to fall and come into contact with a patient or operator.

Product
Captus 4000e Thyroid Uptake System, Model Numbers 5430-30151 and 5430-30152 - Product Usage: is intended to be used by trained nuclear Medicine Technologists or Nuclear Medicine Physicians to perform Thyroid Uptake Procedures.
Manufacturer(s)
Capintec Inc

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov