Home / Recalls / Medical Devices
Capnia Inc recalls Precision Sampling Set, REF: C20206 for use with the CoSense End-Tidal Carbon Monoxide Monitor
Recalled August 6, 2025 · FDA recall Z-2198-2025 · Capnia
Hazard
Sampling set nasal cannula used with End-Tidal Carbon Monoxide (ETCO) monitor may have residual carbon monoxide and volatile organic compounds present due to insufficient aeration of the cannula's carbon filter prior to sealing, which may lead to higher than expected ETOCc scores during clinical evaluations. · FDA Class II
Units
2160
Sold at
US Nationwide distribution in the states of PA, UT, TX, FL, MA, IA.
Description
Sampling set nasal cannula used with End-Tidal Carbon Monoxide (ETCO) monitor may have residual carbon monoxide and volatile organic compounds present due to insufficient aeration of the cannula's carbon filter prior to sealing, which may lead to higher than expected ETOCc scores during clinical evaluations.
Product
Precision Sampling Set, REF: C20206 for use with the CoSense End-Tidal Carbon Monoxide Monitor
Manufacturer(s)
Capnia Inc
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →