Home / Recalls / Medical Devices
Cardinal Health 200, LLC recalls Cardinal Health Jackson-Pratt Channel Drain, 10 Fr, REF JP-2226
Recalled January 25, 2023 · FDA recall Z-0984-2023 · Cardinal Health 200
Hazard
Cardinal Health recently received multiple complaints regarding various performance issues with Cardinal Health Jackson-Pratt Channel Drains including lack of radiopacity in wound drains, dull trocars impacting product performance, lack of flexibility of wound drains, and broken wound tips. · FDA Class II
Units
2277 units
Sold at
Worldwide distribution - US Nationwide and the countries of Canada, Saudi Arabia, Singapore, Hong Kong.
Description
Cardinal Health recently received multiple complaints regarding various performance issues with Cardinal Health Jackson-Pratt Channel Drains including lack of radiopacity in wound drains, dull trocars impacting product performance, lack of flexibility of wound drains, and broken wound tips.
Product
Cardinal Health Jackson-Pratt Channel Drain, 10 Fr, REF JP-2226
Manufacturer(s)
Cardinal Health 200, LLC
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →