RecallWatchUSA

Home / Recalls / Medical Devices

Cardinal Health 200, LLC recalls Cardinal Health MEDI-VAC CRD Suction Canister Liner, 1000 mL Suction Canister Liner Kit with 5 mm 1.8 m Tube, REF 65651

Recalled January 24, 2024 · FDA recall Z-0755-2024 · Cardinal Health 200

Hazard
The above-listed lots of Cardinal Health Medi-Vac CRD" Suction Canister Liners 1L may experience loss of suction on low or low intermittent suction settings, due to a misalignment of the liner with the outer hard canister. · FDA Class II
Units
203244 units
Sold at
US, Canada, Australia and New Zealand

Description

The above-listed lots of Cardinal Health Medi-Vac CRD" Suction Canister Liners 1L may experience loss of suction on low or low intermittent suction settings, due to a misalignment of the liner with the outer hard canister.

Product
Cardinal Health MEDI-VAC CRD Suction Canister Liner, 1000 mL Suction Canister Liner Kit with 5 mm 1.8 m Tube, REF 65651-516
Manufacturer(s)
Cardinal Health 200, LLC

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov