RecallWatchUSA

Home / Recalls / Medical Devices

CareFusion 203, Inc. recalls CareFusion EnVe and ReVel Ventilators. The EnVe/ ReVel Ventilator is designed for use on patients who require respira

Recalled October 1, 2014 · FDA recall Z-2593-2014 · CareFusion 203

Hazard
CareFusion has identified a potential for damage to the power connector on the EnVe and ReVel ventilators due to misconnection by the operator. The input port on the ventilators must have the adapters correctly aligned to function properly. Incorrect alignment may cause the pins in the connector to bend, causing the PTV AC power adapter and/or PTV Auto Lighter Power Cord to be shortened, and pre · FDA Class I
Units
1729
Sold at
Nationwide Distribution including AL, AZ, CA, CO, CT, FL, GA, IL, IN, KS, KY, LA, MA, MI, MN, MS,M0, NE, NV,NH,NH, NY NMC, OH, OR,PA, RID, SC, TN, TX, UT VT, VA, WA, and WI.

Description

CareFusion has identified a potential for damage to the power connector on the EnVe and ReVel ventilators due to misconnection by the operator. The input port on the ventilators must have the adapters correctly aligned to function properly. Incorrect alignment may cause the pins in the connector to bend, causing the PTV AC power adapter and/or PTV Auto Lighter Power Cord to be shortened, and pre

Product
CareFusion EnVe and ReVel Ventilators. The EnVe/ ReVel Ventilator is designed for use on patients who require respiratory support or mechanical ventilation and weigh a minimum of 5 kg (11 lbs.). It is suitable for service in hospital, transport and homecare as a source of continuous or intermittent positive pressure ventilator support, delivered invasively or non-invasively. These ventilators are restricted medical devices intended for operation by a trained personnel under the direction of a physician and in accordance with all applicable state laws and regulations.
Manufacturer(s)
CareFusion 203, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

CareFusion 203, Inc. recalls CareFusion EnVe and ReVel Ventilators. The EnVe/ ReVel Ventilator is designed for use on patients who require respira | RecallWatchUSA