RecallWatchUSA

Home / Recalls / Medical Devices

Carefusion 2200 Inc recalls 1) Adult Manual Resuscitator, 2K8004, Oxygen Reservoir Bag, Adult Mask; 2) Adult Manual Resuscitator, 2K8005, 40" (1.0

Recalled August 21, 2013 · FDA recall Z-1947-2013 · Carefusion 2200

Hazard
CareFusion has received four (4) customer reports of a component (oxygen connector) within the resuscitation bag being occluded which disables the flow of supplemental oxygen to the resuscitation bag and thus may pose a serious patient safety risk. · FDA Class II
Units
14,112 units
Sold at
US Nationwide in the states of AZ, CA, CO, DE, FL, GA, IL, LA, MA, MD, MI, MN, MO, MS, NC, NJ, NM, NY, OH, PA, TX, UT, WA

Description

CareFusion has received four (4) customer reports of a component (oxygen connector) within the resuscitation bag being occluded which disables the flow of supplemental oxygen to the resuscitation bag and thus may pose a serious patient safety risk.

Product
1) Adult Manual Resuscitator, 2K8004, Oxygen Reservoir Bag, Adult Mask; 2) Adult Manual Resuscitator, 2K8005, 40" (1.0 m) Oxygen Reservoir Tubing, Adult Mask; 3) Adult Manual Resuscitator, 2K8017, Variable Volume Oxygen Reservoir Tubing, Adult Mask; 4) Adult Manual Resuscitator, 2K8034, 40" (1.0 m) Oxygen Reservoir Tubing, PEEP Valve, without Mask; 5) Adult Manual Resuscitator, 2K8035, Oxygen Reservoir Bag, PEEP Valve, Adult Mask; 6) Pediatric Manual Resuscitator, 2K8037, 40" (1.0 m) Oxygen Reservoir Tubing, PEEP Valve, Pediatric Mask, Pressure Limiting Valve with Lock; 7) Infant Manual Resuscitator, 2K8040, Oxygen Reservoir Bag, PEEP Valve, Infant Mask, Pressure Limiting Valve with Lock Product Usage: Pulmonary resuscitation
Manufacturer(s)
Carefusion 2200 Inc

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Carefusion 2200 Inc recalls 1) Adult Manual Resuscitator, 2K8004, Oxygen Reservoir Bag, Adult Mask; 2) Adult Manual Resuscitator, 2K8005, 40" (1.0 | RecallWatchUSA