RecallWatchUSA

Home / Recalls / Medical Devices

Carefusion 2200 Inc recalls Safe-T-Centesis PLUS 8 Fr Catheter Drainage Tray

Recalled July 29, 2020 · FDA recall Z-2652-2020 · Carefusion 2200

Hazard
The Universal Drainage Set (UDS) may contain valves that leak internally which may result in the introduction of up to 30cc of air into the patient. · FDA Class II
Units
N/A
Sold at
370 US customers and 13 distributors. Product distributed in: AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MD, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY Customer listed, including Federal distribution, uploaded.

Description

The Universal Drainage Set (UDS) may contain valves that leak internally which may result in the introduction of up to 30cc of air into the patient.

Product
Safe-T-Centesis PLUS 8 Fr Catheter Drainage Tray
Manufacturer(s)
Carefusion 2200 Inc

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Carefusion 2200 Inc recalls Safe-T-Centesis PLUS 8 Fr Catheter Drainage Tray | RecallWatchUSA