Home / Recalls / Medical Devices
CareFusion 303, Inc. recalls Alaris PC unit, model 8015, with version 9.12-Keyboard Processor and logic board assembly part number TC10006584. The
Recalled August 14, 2013 · FDA recall Z-1835-2013 · CareFusion 303
Hazard
CareFusion is recalling the Alaris PC units model 8015 (PC unit), version 9.12, because it is operating at an incorrect voltage and it could potentially experience a loss of communication between the PC Unit main Processor and Keyboard Processor. · FDA Class I
Units
26,564 total units (26,220 units for US)
Sold at
Worldwide distribution: US (nationwide) and countries of: Australia, Canada and Europe.
Description
CareFusion is recalling the Alaris PC units model 8015 (PC unit), version 9.12, because it is operating at an incorrect voltage and it could potentially experience a loss of communication between the PC Unit main Processor and Keyboard Processor.
Product
Alaris PC unit, model 8015, with version 9.12-Keyboard Processor and logic board assembly part number TC10006584. The Alaris PC Unit is the main user interface unit and power supply of the Alaris System, a modular system to be used with Alaris System modules intended for use in todays growing professional healthcare environment for facilities that utilize infusion and/or monitoring devices.
Manufacturer(s)
CareFusion 303, Inc.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →