CareFusion 303, Inc. recalls Alaris Pump, Model 8100, with bezel mechanical assemblies manufactured with FR-110 plastic resin. The following produ

Recalled July 24, 2019 · FDA recall Z-1768-2019 · CareFusion 303

Hazard
The bezel manufacturing process for the FR-110 plastic may have resulted in its weakening. A bezel with weakened plastic may, over time, lead to separation of the bezel post (recall issue) as well as other damage to the bezel (i.e. external cracking). The separation of one or more bezel posts may result in free flow, over infusion, under infusion or interruption of infusion. · FDA Class I
Units
607618
Sold at
Worldwide Distribution - U.S. Nationwide CA, TX, NC, AL, ID, TN, IN, MA, UT, HI, NH, PA, IL, OR, FL, MI, VA, AR, OH, GA, NY, MD, MN, NM, AZ, KY, WI, NE, MS, DC, MO, KS, SC, LA, MT, WA, NJ, CO, CT, OK, AK, AS, CU, DE, GU, IA, ME, MP, ND, NV, PR, SD, VT, WV, WY; OUS: AE, AU, BH, CA, CH, CN, GI, IL, IN, KW, MX, MY, NL, NZ, PH, PK, QA, SA, SG, TW, and ZA.

Description

The bezel manufacturing process for the FR-110 plastic may have resulted in its weakening. A bezel with weakened plastic may, over time, lead to separation of the bezel post (recall issue) as well as other damage to the bezel (i.e. external cracking). The separation of one or more bezel posts may result in free flow, over infusion, under infusion or interruption of infusion.

Product
Alaris Pump, Model 8100, with bezel mechanical assemblies manufactured with FR-110 plastic resin. The following products are affected by this recall as they are pumps or pump assemblies with bezels manufactured between April 2011 and June 2017 with the FR-110 plastic. Alaris Pump Module serviced with LVP Mechanism Sub Assembly (P/N 10942012, P/N 49000007, and P/N 49000203), and Alaris Pump module Bezel Kit Assembly (P/N 10964559 and P/N 49000204). Product Usage: The Alaris Pump module is a large volume infusion pump offered under the Alaris System. The Alaris Pump module will deliver medication and fluids using the IV administration sets for continuous or intermittent delivery through clinically acceptable routes of administration such as intravenous or epidural. There are 6 bezel posts on the Alaris Pump bezel assembly.
Manufacturer(s)
CareFusion 303, Inc.