RecallWatchUSA

Home / Recalls / Medical Devices

CareFusion 303, Inc. recalls BD Pyxis: MedStation ES (Main), REF: 323; MedStation ES Tower, REF: 352; Anesthesia Station ES, REF: 327; CII Safe ES

Recalled July 24, 2024 · FDA recall Z-2374-2024 · CareFusion 303

Hazard
If automated dispensing cabinets have specific software versions, and Component Manager was configured to "installed mode", and software patch KB 5033688 was installed in the production environment, then this will will lead to an error that prevents the user from accessing the dispensing software application, which may delay accessing medication. · FDA Class II
Units
326
Sold at
Worldwide - US Nationwide distribution including in the states of OR, KS, NE, LA, ND, OK, AL, AK, WI, WV, WA, NV, PR, TX, CO, OH, CA, MO, IA, AZ, IN, AR, IL, MN, HI, NM, FL, SD, UT, NY, VA, GA, MA, NC and the countries of TW, CA.

Description

If automated dispensing cabinets have specific software versions, and Component Manager was configured to "installed mode", and software patch KB 5033688 was installed in the production environment, then this will will lead to an error that prevents the user from accessing the dispensing software application, which may delay accessing medication.

Product
BD Pyxis: MedStation ES (Main), REF: 323; MedStation ES Tower, REF: 352; Anesthesia Station ES, REF: 327; CII Safe ES, REF: 1116-00; CII Safe ES, Desktop PC, Wired Scanner, REF: 107-254-01; CII Safe ES, Desktop PC, Wireless Scanner, REF: 107-255-01; CII Safe ES, Desktop PC, Wired Scanner 1.7.3, REF: 107-256-01; CII Safe ES, Desktop PC, Wireless Scanner 1.7.3, REF: 107-257-01
Manufacturer(s)
CareFusion 303, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov