Home / Recalls / Medical Devices
CAREFUSION recalls BD Alaris Pump Infusion Buerette Set Smallbore Tubing Smartsite Port (Burette) 3 SmartSite Y-Sites, REF: 2441-0006
Recalled September 22, 2021 · FDA recall Z-2466-2021 · CAREFUSION
Hazard
Needle-free connector valves, either standalone or attached to extension sets, may have issues including difficulty to flush, flow issues, and partial or total occlusions, which could lead to therapy delay. Follow-up activates not required if product has already been used or was safely connected to the female luer of connecting device. · FDA Class II
Units
270
Sold at
Worldwide distribution - US Nationwide distribution in the states of FL, PA, NC, OH, SC, IL, RI, MO, NY, TN, AL, VA, PR, CA, NE, NJ, MI, MD, NM, LA, KY, NV, ME, OK, WA, MA, WV, TX, AR, OR, KS, GA, VT, IN, WY, AZ, CO, MN, WI, UT, SD, MS, ID, MT, ND, HI, IA, DC, AK, NH, CT, DE, GU and the countries of CA, NZ, TW, AU, BE, VN, AR, MY, TH, KR, ID, PH, SG, BH, SA, HK, IN, CN, BR.
Description
Needle-free connector valves, either standalone or attached to extension sets, may have issues including difficulty to flush, flow issues, and partial or total occlusions, which could lead to therapy delay. Follow-up activates not required if product has already been used or was safely connected to the female luer of connecting device.
Product
BD Alaris Pump Infusion Buerette Set Smallbore Tubing Smartsite Port (Burette) 3 SmartSite Y-Sites, REF: 2441-0006
Manufacturer(s)
CAREFUSION
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →