RecallWatchUSA

Home / Recalls / Medical Devices

Carestream Health, Inc. recalls DRX-Compass/DR-FIT X-ray Systems with Firmware version 2.3.2.0/2.3.2.4. For use in obtaining diagnostic quality radiogr

Recalled March 8, 2023 · FDA recall Z-1206-2023 · Carestream Health

Hazard
After pressing and releasing the Z-Axis Motorized buttons on the Tube Head Display, the Overhead Tube Crane (OTC) could potentially continue the movement unexpectedly to a certain distance which could potentially result in a injury · FDA Class II
Units
72 units
Sold at
Worldwide distribution - US Nationwide distribution in the states of California, Idaho, Illinois, Massachusetts, Minnesota, Nevada, North Carolina, Texas Wisconsin and the countries of Argentina, Australia, Brazil, China, Finland, France, INDIA, ITALY, Korea, Malaysia, Poland, Republic of Korea.

Description

After pressing and releasing the Z-Axis Motorized buttons on the Tube Head Display, the Overhead Tube Crane (OTC) could potentially continue the movement unexpectedly to a certain distance which could potentially result in a injury

Product
DRX-Compass/DR-FIT X-ray Systems with Firmware version 2.3.2.0/2.3.2.4. For use in obtaining diagnostic quality radiographic images to aid the physician with diagnosis. CATALOG #(s): 8624181 X-RAY SYSTEM CONSOLE PDU 3P 380-480 C/F 8624199 X-RAY SYSTEM CONSOLE PDU 3P 208-240 C/F 8621617 X-RAY SYSTEM CONSOLE PDU 3P 380-480 C/F 8622359 X-RAY SYSTEM CONSOLE PDU 3P 208-240 C/F Note: PDU is a standalone Power Distribution Unit (PDU), plugged into the hospital outlet, which supplies Alternate Current (AC) to the Overhead Tube Crane (OTC).
Manufacturer(s)
Carestream Health, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Carestream Health, Inc. recalls DRX-Compass/DR-FIT X-ray Systems with Firmware version 2.3.2.0/2.3.2.4. For use in obtaining diagnostic quality radiogr | RecallWatchUSA