RecallWatchUSA

Home / Recalls / Medical Devices

Carl Zeiss Metrology Inc recalls Metrotom 800 (130kV CT scanner) Industrial CT scanner for measuring and inspecting complete components made of plasti

Recalled February 7, 2018 · FDA recall Z-0379-2018 · Carl Zeiss Metrology

Hazard
Due to a software error, it was observed that when the user closes the access door, the system will resume its scan cycle automatically, instead of the user initiating x-ray generation from the control panel. · FDA Class II
Units
13 US
Sold at
Worldwide Distribution - US, including the states of : WI, NC, GA, IL, AZ, FL, CA, NJ, and territory of PR; and the country of Canada.

Description

Due to a software error, it was observed that when the user closes the access door, the system will resume its scan cycle automatically, instead of the user initiating x-ray generation from the control panel.

Product
Metrotom 800 (130kV CT scanner) Industrial CT scanner for measuring and inspecting complete components made of plastic or light metal.
Manufacturer(s)
Carl Zeiss Metrology Inc

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov