Home / Recalls / Medical Devices
Carl Zeiss Suzhou Co., Ltd. recalls OPMI LUMERA 300, REF 6137
Recalled November 1, 2023 · FDA recall Z-0183-2024 · Carl Zeiss Suzhou
Hazard
Surgical microscope for use in ophthalmology may be missing a screw from the suspension arm, which may cause the suspended components to fall from the suspension arm, which may injure any person under the suspended components. · FDA Class II
Units
8
Sold at
US Nationwide distribution in the states of MD, TX, CA, GA, LA.
Description
Surgical microscope for use in ophthalmology may be missing a screw from the suspension arm, which may cause the suspended components to fall from the suspension arm, which may injure any person under the suspended components.
Product
OPMI LUMERA 300, REF 6137
Manufacturer(s)
Carl Zeiss Suzhou Co., Ltd.
Made in
China
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →