RecallWatchUSA

Home / Recalls / Medical Devices

Cellavision AB recalls REF: CC286297 or BS737162 Automated Digital Cell Morphology analyzer DI-60, IVD in conjunction with a Honeywell Vuquest

Recalled November 26, 2025 · FDA recall Z-0542-2026 · Cellavision AB

Hazard
Automated cell-locating device barcode reader may read the barcode of the previously processed slide resulting in a misattribution of diagnostic results. · FDA Class II
Units
688 units
Sold at
US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, VT, WA, WI, WV, WY, PR.

Description

Automated cell-locating device barcode reader may read the barcode of the previously processed slide resulting in a misattribution of diagnostic results.

Product
REF: CC286297 or BS737162 Automated Digital Cell Morphology analyzer DI-60, IVD in conjunction with a Honeywell Vuquest or Jadk FM-5 barcode reader component.
Manufacturer(s)
Cellavision AB
Made in
Sweden

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov