Centurion Medical Products Corporation recalls Centurion CIRCLAMP W/1.1CM BELL - single-use, disposable instrument used to compress the foreskin of the penis during c

Recalled October 16, 2019 · FDA recall Z-0120-2020 · Centurion Medical Products

Hazard
A mismatch between the angle of the hole in the base and the angle of the bell, which may cause the clamp to remove foreskin without the use of a scalpel. · FDA Class II
Units
1530 each
Sold at
Nationwide

Description

A mismatch between the angle of the hole in the base and the angle of the bell, which may cause the clamp to remove foreskin without the use of a scalpel.

Product
Centurion CIRCLAMP W/1.1CM BELL - single-use, disposable instrument used to compress the foreskin of the penis during circumcision of a male infant. ReOrder: 310CR
Manufacturer(s)
Centurion Medical Products Corporation

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov