RecallWatchUSA

Home / Recalls / Medical Devices

CHG Hospital Beds Inc recalls Spirit Plus Bed, A-C Powered Hospital Bed

Recalled August 10, 2016 · FDA recall Z-2331-2016 · CHG Hospital Beds

Hazard
Stryker Medical is initiating a voluntary recall of the Spirit Select and Spirit Plus A-C powered hospital beds due to reports of hi-lo actuators broken at the mount ends, which could cause the bed to unexpectedly lower resulting in patient injury. · FDA Class II
Units
4,308 units total
Sold at
Worldwide Distribution: US (nationwide) including states of: AK , AL, AR, AZ, CA, CO, CT, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, VA, WA, and WI; and country of: Canada

Description

Stryker Medical is initiating a voluntary recall of the Spirit Select and Spirit Plus A-C powered hospital beds due to reports of hi-lo actuators broken at the mount ends, which could cause the bed to unexpectedly lower resulting in patient injury.

Product
Spirit Plus Bed, A-C Powered Hospital Bed
Manufacturer(s)
CHG Hospital Beds Inc
Made in
Canada

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov