Childrens Medical Ventures recalls Philips/Children's Medical Ventures Smart Monitor 2 Professional Series infant apnea monitor.

Recalled June 25, 2014 · FDA recall Z-1789-2014 · Childrens Medical Ventures

Hazard
The battery pack wire harness is improperly assembled rendering the device inoperable. The wiring issue causes an error code to be displayed and results in the device sounding an alarm repeatedly at power up. The presence of the error code and the devices alarm prevents the device from functioning until the issue is addressed. · FDA Class II
Units
112
Sold at
Worldwide Distribution - USA including the states of MI, WI, SC, FL, TX, KS, MA, AZ, UT, WV, IN, MN, NV and VA., and the countries of Australia, Belgium and Canada.

Description

The battery pack wire harness is improperly assembled rendering the device inoperable. The wiring issue causes an error code to be displayed and results in the device sounding an alarm repeatedly at power up. The presence of the error code and the devices alarm prevents the device from functioning until the issue is addressed.

Product
Philips/Children's Medical Ventures Smart Monitor 2 Professional Series infant apnea monitor.
Manufacturer(s)
Childrens Medical Ventures