Civco Medical Instruments Co. Inc. recalls eTRAX Needle System Starter Kit 16G(for Aurora Trackers), Part Number 667-151

Recalled April 8, 2026 · FDA recall Z-1733-2026 · Civco Medical Instruments

Hazard
There was an error in inspection and programming of the eTRAX needle sensor for Aurora trackers. The result is a potential for the needle tip position to be incorrectly identified on the user interface. · FDA Class II
Units
18 units
Sold at
US States: OH, PA, WA. China.

Description

There was an error in inspection and programming of the eTRAX needle sensor for Aurora trackers. The result is a potential for the needle tip position to be incorrectly identified on the user interface.

Product
eTRAX Needle System Starter Kit 16G(for Aurora Trackers), Part Number 667-151
Manufacturer(s)
Civco Medical Instruments Co. Inc.