RecallWatchUSA

Home / Recalls / Medical Devices

Claymount Assembli recalls The Claymount Harmony is a grid cabinet designed to accept a digital x-ray imaging receptor and stationary grid. It cont

Recalled July 9, 2014 · FDA recall Z-1836-2014 · Claymount Assembli

Hazard
Claymount Assemblies BV is recalling the Harmony2 model of Harmony Bucky because 1) the devices are incorrectly labeled in that the label does not list the manufacturer address, the date of manufacture, or a statement of compliance with 21 CFR Subchapter J, and 2) the device's insulation values do not comply with IEC 60601-1 in all cases. · FDA Class II
Units
12 devices
Sold at
Worldwide Distribution - USA including NC, OH, CO, CA, NJ, IL, OH, PA, FL and Internationally to Canada, China, France, Italy, Netherlands, South Africa, Switzerland, Turkey, and United Kingdom.

Description

Claymount Assemblies BV is recalling the Harmony2 model of Harmony Bucky because 1) the devices are incorrectly labeled in that the label does not list the manufacturer address, the date of manufacture, or a statement of compliance with 21 CFR Subchapter J, and 2) the device's insulation values do not comply with IEC 60601-1 in all cases.

Product
The Claymount Harmony is a grid cabinet designed to accept a digital x-ray imaging receptor and stationary grid. It contains signaling electronics to indicate the presence of the grid and detector as well as detector orientation. It may be mounted to a wall stand or incorporated in a radiographic table. It is intended to be used in general radiographic applications as a component of a radiographic system.
Manufacturer(s)
Claymount Assembli
Made in
Netherlands

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Claymount Assembli recalls The Claymount Harmony is a grid cabinet designed to accept a digital x-ray imaging receptor and stationary grid. It cont | RecallWatchUSA