RecallWatchUSA

Home / Recalls / Medical Devices

Codan US Corporation recalls CODAN SWAN-LOCK Swabable Needlefree Connector, Catalog #BC1000N, REF 16.7220, individually packaged, 504/case, Sterile.

Recalled July 8, 2020 · FDA recall Z-2433-2020 · Codan US

Hazard
Individual packages of connectors may have incomplete or bad seals which would compromise sterility. · FDA Class II
Units
34,776 units
Sold at
Distribution was made to CA, MI, NY, PA, and VA. There was no military/government distribution. Foreign distribution was made to Canada, Australia, New Zealand, and Chile.

Description

Individual packages of connectors may have incomplete or bad seals which would compromise sterility.

Product
CODAN SWAN-LOCK Swabable Needlefree Connector, Catalog #BC1000N, REF 16.7220, individually packaged, 504/case, Sterile. The firm name on the label is CODAN US CORPORATION, Santa Ana, CA.
Manufacturer(s)
Codan US Corporation

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Codan US Corporation recalls CODAN SWAN-LOCK Swabable Needlefree Connector, Catalog #BC1000N, REF 16.7220, individually packaged, 504/case, Sterile. | RecallWatchUSA