RecallWatchUSA

Home / Recalls / Medical Devices

Comet Technologies Usa Inc recalls YXLON Cougar Cabinet X-Ray System Product Usage: The products are generally used for non-destructive testing and qua

Recalled August 14, 2019 · FDA recall Z-2106-2019 · Comet Technologies Usa

Hazard
Failure to meet the requirement of the performance standard for cabinet x-ray systems (21 CFR 1020.40) for a door interlock that causes physical disconnection of the power supply using only the movement of the door. · FDA Class II
Units
93
Sold at
US Nationwide Distribution

Description

Failure to meet the requirement of the performance standard for cabinet x-ray systems (21 CFR 1020.40) for a door interlock that causes physical disconnection of the power supply using only the movement of the door.

Product
YXLON Cougar Cabinet X-Ray System Product Usage: The products are generally used for non-destructive testing and quality control
Manufacturer(s)
Comet Technologies Usa Inc

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov