RecallWatchUSA

Home / Recalls / Medical Devices

Companion Medical, Inc. recalls InPen Diabetes Management App, part of the InPen System, REF: MMT-105ELBLNA, MMT-105ELGYNA, MMT-105ELPKNA, MMT-105ELPKNA

Recalled December 28, 2022 · FDA recall Z-0741-2023 · Companion Medical

Hazard
Diabetes management app may crash immediately upon opening, while using the app and/or while app is running in the background, which may lead to inability to use the app. An error message may be displayed indicating the app has stopped or isn't responding. After the crash, a present dose reminder is not displayed, which may lead to delayed insulin therapy and potentially resulting in hyperglycemia. · FDA Class II
Units
140
Sold at
US Nationwide distribution including in the states of CA, NY, MT, OH, IN, MN, PA, CT, TX, WA, TN, SD, MI, FL, OR, NC, KY, WI, MO, WV, NV, SC, AZ, AR, MA, LA, VT, GA.

Description

Diabetes management app may crash immediately upon opening, while using the app and/or while app is running in the background, which may lead to inability to use the app. An error message may be displayed indicating the app has stopped or isn't responding. After the crash, a present dose reminder is not displayed, which may lead to delayed insulin therapy and potentially resulting in hyperglycemia.

Product
InPen Diabetes Management App, part of the InPen System, REF: MMT-105ELBLNA, MMT-105ELGYNA, MMT-105ELPKNA, MMT-105ELPKNA, MMT-105NNBLNA, MMT-105NNGYNA, MMT-105NNPKNA.
Manufacturer(s)
Companion Medical, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Companion Medical, Inc. recalls InPen Diabetes Management App, part of the InPen System, REF: MMT-105ELBLNA, MMT-105ELGYNA, MMT-105ELPKNA, MMT-105ELPKNA | RecallWatchUSA