RecallWatchUSA

Home / Recalls / Medical Devices

Compass Health Brands (Corporate Office) recalls CPAP Mask Cushion, Model Nos. PB781S, PB781M, PB781L, 781S, 781M and 781L Product Usage: A CPAP mask is attached to

Recalled August 29, 2018 · FDA recall Z-2741-2018 · Compass Health Brands (Corporate Office)

Hazard
A May 2015 design change in the cushion seal replacement part and accompanying elbow replacement part resulted in incompatibility of the new cushion seal with the previous design of the elbow. Use of the new cushion seal with the previous design elbow could lead to unacceptable CO2 porting. As there was no model number change for this design change, customers may not be aware of the incompatibility. · FDA Class I
Units
742
Sold at
US nationwide distribution to AL, CT, FL, GA, KY, MS, NC, NY, TN, and TX.

Description

A May 2015 design change in the cushion seal replacement part and accompanying elbow replacement part resulted in incompatibility of the new cushion seal with the previous design of the elbow. Use of the new cushion seal with the previous design elbow could lead to unacceptable CO2 porting. As there was no model number change for this design change, customers may not be aware of the incompatibility.

Product
CPAP Mask Cushion, Model Nos. PB781S, PB781M, PB781L, 781S, 781M and 781L Product Usage: A CPAP mask is attached to an end user s face and a CPAP machine, to treat end users with sleep apnea. The replacement cushion is a non-durable component that requires replacement every 3 - 6 months. The cushion is used for both comfort and to provide a seal between the face and the mask.
Manufacturer(s)
Compass Health Brands (Corporate Office)

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Compass Health Brands (Corporate Office) recalls CPAP Mask Cushion, Model Nos. PB781S, PB781M, PB781L, 781S, 781M and 781L Product Usage: A CPAP mask is attached to | RecallWatchUSA